According to
EN 1822-4, IS0 29463-4, EN 1822-5 ISO 29463-5, GB/T 6165, etc
Test indexes
Filter efficiency
Pressure drop
Air volume vs. resistance(pressure drop) curve
Leak detection by scan method
Advantages
· Automated Workflow: Reduces manual operations and improves efficiency.
· Real-Time Display: Automatically records, calculates, and analyzes test results.
· Smart Alerts: Built-in notification system minimizes human errors.
· User-Friendly Interface: Intuitive and easy to use with minimal training required.
· Flexible Report Export: Generates editable Excel reports and printable labels.
· Data Management & Backup: Supports automatic data storage and backup to prevent data loss.
· Remote Access: Enables remote diagnostics, debugging, and calibration.
Software and Results
PLC and computer control data collection anddevice actions, with a data collection time ofmilliseconds. The computer analyzes the dataand displays segmented two-dimensional images on LCD screen, real-time observation of the entire scanning and testing process.
A4 report will generated automatically afterthe test, and all information will reflect in thereport. 3D image can be generated.
Configure a label printer, test results can beprinted as adhesive label, pasted on the filter.
Technical Data
Flow rate: 500~4500 m3/h
Differentia pressure: Range 0~800 Pa, Accuracy 士1% F·S
Filter dimensions: Max.1260x1260x400mm
Min.150x150x50mm(1 probe)
350x150x50mm(4 probe)
Filter classes: H10~U17, 85%~99.9999995%@MPPS
Test aerosol: DEHS
Particle counter: 8 channels, 0.1,0.15,0.2,0.25,0.3,0.5,0.7,1.0um
Scan speed: <10 cm/s
Sampling probes: 1 or 4 probes, 66X15mm
Compressed air supply: >7 kg/cm2
Power supply: AC 380V 50Hz
Dimensions: 5900x2000x2250mm