English
简体中文
繁體中文
العربية
Pусский
Español
한국어
Türk dili

Home » Knowledge Center » Expert Ideas » Standard-ISO 23328 Breathing System Filters for Anaesthetic and Respiratory Use

Standard-ISO 23328 Breathing System Filters for Anaesthetic and Respiratory Use

Publish Time: 2021-12-23     Origin: Site

ISO 23328-1:2008 Breathing System Filters for Anaesthetic and Respiratory Use- Part 1: Salt-test method to assess filtration performance

This part of ISO 23328 gives a method of test for assessing the filtration performance of breathing system filters (BSF).

In the test, the BSF is challenged with sodium chloride particles of the most penetrating size range, 0.1~0.3μm.


Apparatus

Flowmeter, with an accuracy of ±5% of the actual value to be measured,

Sodium chloride aerosol generator, capable of generating an aerosol at (25±5)℃ and relative humidity of (30±10)% with a concentration between 10mg/m3 and 20mg/m3 which has been neutralized to the Boltzmann equilibrium state.

Scanning mobility particle sizer, or equivalent instrument.

Suitable forward-light-scattering photometer, or equivalent instrument.


Flowrates for testing BSF

Paediatric: 15L/min

Adult: 30L/min


If you have relevant testing requirements, we recommend our SC-FT-1406D-Plus Automatic Filter Tester to you.

Flowrate: 0~99.9 L/min

Pressure gauge: 0~125Pa, 0~500Pa or customized

Test aerosol: NaCl and paraffin oil

Accuracy of flowmeter: 2.5%

Accuracy of pressure gauge: 1%


Contact Us