Views: 0 Author: Purgate Purgate auditum tempus: 2021-12-23 Origin: Situs
ISO 23328-1: MMVIII Heading Ratio Filtra ad Papaver et Respiratoriorum Use- Part I: Salis-Test modum ad assess filtration perficientur
Haec parte ISO (XXIII) CCCXXVIII dat modum test ad perpendendis filtration perficientur de spirandi ratio Filtra (BSF).
In test, in BSF provocatur cum sodium chloride particulas maxime penetrantes magnitudine rhoncus, 0.1 ~ 0.3μm.
Apparatus
Flowmeter, cum accurate de ± V% de ipsa valorem ut metiri:
Sodium chloride Aerosol generator, capax generandi ad Aerosol ad (XXV ± V) ℃ et relative humiditas (XXX ± X)% cum concentratione inter 10mg / M 3 et 20mg / m 3 , qui est neutralized ad Boltzmann aequilibrium statu.
Scanning mobilitatem particula sizer, aut equivalent instrumentum.
Idoneam deinceps-lux-spargere photometer vel equivalent instrumentum.
Floures ad temptationis BSF
Pediatric: 15l / min
Adult: 30L / min
Si vos have relevant temptationis requisitis, nos suadeo nostrum SC-ft-1406D-plus automatic filter testem vobis.
Florat: 0 ~ 99.9 l / min
PRESSIO MODULUS: 0 ~ 125PA, 0 ~ 500PA et Nativus
Test Aerosol: Nacl et Paraffin oleum
Accuratio of Flowmeter: 2.5%
Accurate pressura MODULUS: I%