How to Choose HEPA Filter Testing Equipment

共有:

目次

HEPA testing equipment is not one type of machine. Media development, finished-filter efficiency, factory scan leak testing and installed-filter integrity verification involve different test objects, outputs and instruments. The correct system can be selected only after the test question is clearly defined.

Four test tasks that must be separated

Four test tasks that must be separated

Test Taskテスト項目Primary OutputTypical Equipment
Media R&DFlat-sheet mediaFractional efficiency, 圧力降下, MPPSFilter media tester
Integral efficiencyComplete filterOverall efficiency or penetration and pressure dropFull-size duct test bench
Factory scanFilter face and critical seal areasLocal penetration and leak locationAutomatic scan leak test system
Installed integrityInstalled filter and installation boundaryIntegrity of filter, seals and housing in the fieldPortable aerosol generator with photometer or particle-counter system

How do standards and equipment correspond?

How do standards and equipment correspond?

ISO 29463-3 applies to flat-sheet high-efficiency filter media. ISO 29463-5 covers efficiency testing of complete filter elements near MPPS, while the scan method in ISO 29463-4 is used to locate local leakage. After filters are installed in a cleanroom or controlled environment, the objective changes to verification of the installed filtration system. This is normally performed under ISO 14644-3, a project specification or an industry procedure. A field result does not replace the manufacturer’s classification, and a factory scan cannot prove that the installed system has no bypass leakage.

When is a fixed 0.3 µm test useful?

When is a fixed 0.3 µm test useful?

A fixed-particle-size efficiency or leak check can be useful for fast production quality control. しかし, if the applicable standard requires testing at MPPS or scanning against local penetration limits, ある 0.3 µm result alone cannot demonstrate full compliance with ISO 29463 または EN 1822. A rapid internal method should be correlated with the standard method, and its purpose should be stated clearly in the report.

Six questions to answer before choosing a system

Six questions to answer before choosing a system

  • Are you testing flat-sheet media or a complete filter?
  • Do you need integral efficiency, local scanning or installed integrity verification?
  • What are the product dimensions, construction, rated airflow and maximum resistance?
  • What target class and edition of the standard apply?
  • Which aerosol and detector will be used, and is the expected downstream concentration measurable?
  • What throughput, オートメーション, reporting and metrological traceability are required?

Where do SCPUR systems fit?

Where do SCPUR systems fit?

The SC-FT-1406 series is used for efficiency and pressure-drop evaluation of flat-sheet media. の SC-L8023 そして SC-L8023U are designed for integral efficiency and automatic scan leak testing of complete high-efficiency filters. Installed-filter integrity verification requires portable field instruments and a procedure matched to the project. Providing the filter design, applicable clauses and expected report before quotation is the most reliable way to obtain the right configuration instead of a generic HEPA test bench.

EDITORIAL REFERENCES

Primary Standards Checked

Primary Standards Checked

The following official ISO pages were checked for this rewrite. Formal testing, contracts and compliance decisions should be based on the purchased standard text and the edition specified by the project.

  • ISO 16890-1:2016Technical specifications, requirements and classification system based on particulate matter efficiency for general-ventilation air filters
  • ISO 16890-2:2022Measurement of fractional efficiency and airflow resistance
  • ISO 16890-3:2024Gravimetric efficiency and airflow resistance versus captured test-dust mass
  • ISO 16890-4:2022Conditioning method used to determine minimum fractional test efficiency
  • ISO 29463-1:2024 – 分類, performance testing and marking of high-efficiency filters and media
  • ISO 29463-3:2011Testing flat-sheet filter media
  • ISO 29463-4:2011Scan method for determining leakage of filter elements; confirmed in 2024 while a new edition is under development
  • ISO 29463-5:2022 – MPPS efficiency test method for filter elements
  • ISO 14644-3:2019Test methods for cleanrooms and clean zones, including performance verification relevant to installed systems

連絡する

出発前に...

自分のプロジェクトにどの国際標準が適用されるかわからない? 当社の 10 年間の業界専門知識に基づいた個別の技術評価については、当社の上級エンジニアと相談してください。.

*私たちはあなたの秘密を尊重し、すべての情報が保護されます.

世界標準の試験システムで品質を検証

当社の専門家がお客様に合わせてカスタム ソリューションを設計いたします. あなたのプロジェクトについて教えてください, あなたの品質を正確かつ迅速に検証しましょう.

カスタムR&D & カスタマイズされたソリューション

カスタムR&D & カスタマイズされたソリューション

世界標準への準拠

世界標準への準拠

100% 自社開発のソフトウェアシステム

100% 自社開発ソフトウェアシステム

リモート校正 & グローバルサポート

リモート校正 & グローバルサポート

お問い合わせ

*私たちはあなたの秘密を尊重し、すべての情報が保護されます.

This page has been automatically translated from the English original for convenience. Product models, standards, numerical values and technical limits should be verified against the English version. Please contact SCPUR for project-specific confirmation.