SC-1705
Functions
The SC-1705 is mainly used to test the air tightness of various full masks and half masks, and is suitable for the quality control of mask by producers and users.
According to
EN 136 Section 8.13 Leaktightness
Advantages
The equipment is equipped with a vacuum pump to ensure that the standard requires negative pressure.
The instrument can run automatically, and the operator needs a small amount of work to perform a test. Place the mask over the matching test head, seal the inhalation valve and wet the exhalation valve. Then click the start button, the program run and complete automatically.
The touch screen can intuitively display test results, system status information and operating parameters, greatly reducing the amount of training for users and improving detection efficiency.
Built-in printer and the result will be printed automatically after the test.
The instrument has simple operation and low maintenance cost.
Technical parameters
1. Vacuum Pump
Pumping rate: 2L/min
2. Flow
Range: 0-1000ml/min
Accuracy:3%
3. Differential Pressure
Range: 0-1200Pa
Accuracy: 1Pa
4. Volume of chamber: 150ml
5. Power Supply: AC220V 50Hz
6. Power: 50W
7. Dimension: 340mm*270mm*150mm(Without head mold)
Environment requirements
1. Temperature: (23±5)℃
2. Humidity: 20-75%
Functions
The SC-1705 is mainly used to test the air tightness of various full masks and half masks, and is suitable for the quality control of mask by producers and users.
According to
EN 136 Section 8.13 Leaktightness
Advantages
The equipment is equipped with a vacuum pump to ensure that the standard requires negative pressure.
The instrument can run automatically, and the operator needs a small amount of work to perform a test. Place the mask over the matching test head, seal the inhalation valve and wet the exhalation valve. Then click the start button, the program run and complete automatically.
The touch screen can intuitively display test results, system status information and operating parameters, greatly reducing the amount of training for users and improving detection efficiency.
Built-in printer and the result will be printed automatically after the test.
The instrument has simple operation and low maintenance cost.
Technical parameters
1. Vacuum Pump
Pumping rate: 2L/min
2. Flow
Range: 0-1000ml/min
Accuracy:3%
3. Differential Pressure
Range: 0-1200Pa
Accuracy: 1Pa
4. Volume of chamber: 150ml
5. Power Supply: AC220V 50Hz
6. Power: 50W
7. Dimension: 340mm*270mm*150mm(Without head mold)
Environment requirements
1. Temperature: (23±5)℃
2. Humidity: 20-75%