Send us
           bank@scpur.com
    Whatsapp
 +86 17685707658
 
Home » Products » Other » Mask Breathing Testing Machine According to EN149 SC-RT-1703EN

Product Category

loading

Mask Breathing Testing Machine According to EN149 SC-RT-1703EN

This instrument is used to test exhalation and inhalation resistance of masks. According to the different standards, the models are divided into SC-RT-1703GB, SC-RT-1703KR and SC-RT-1703EN.
Availability:
Quantity:
facebook sharing button
twitter sharing button
line sharing button
wechat sharing button
linkedin sharing button
pinterest sharing button
sharethis sharing button
  • SC-RT-1703EN

  • scince


Review

Used to test exhalation and inhalation resistance for mask.

Comply with the test requirements of EN149 regarding exhalation resistance with 160L/min flow .

Under rapid test mode, there are two kinds of flow for innalation/exhalation; 30/160 (L/min) and 95/160(L/min).

Equipped with adult head model under Europe standard.

Equipped with a headform bracket, it can measure the breathing resistance in five directions.

Controlled by microcomputer, touch screen is easy to operate

Help the head model to wear mask, the program will run automatically after starting test, and the whole test procedure will be             completed automatically.

Clearly display data, build—in printer, the result can be printed directly.



Applications

Used to test exhalation and inhalation resistance for the mask with FFP1/FFP2/FFP3 grade.


Technical Parameters

Test Flow for   Exhalation

160L/Min

Test Flow for   Inhalation

30L/Min and   95L/Min

Flow Rate   Accuracy

±2.5%

Resistance   Range

0~500PA

Differential   Pressure Gauge                Accuracy

±1%

Power

AC220V, 50 or   60 Hz,  Customized Available