ISO 16890 試験装置: What Does a Complete System Include?

共有:

目次

「ISO 16890 test equipment” does not refer to one instrument.

A complete laboratory can include fractional-efficiency measurement, 通気抵抗, filter conditioning, 粉塵積載, weighing and data-processing functions. These functions may use connected modules, but they do not all perform the same test.

The correct configuration depends on whether the laboratory needs product development, routine quality control or complete testing according to the ISO 16890 シリーズ.

Define the Test Scope First

機器を選ぶ前に, determine which results are required:

  • Initial fractional-efficiency curve
  • Conditioned fractional efficiency
  • ISO ePM classification
  • Airflow–resistance curve
  • 重量効率
  • 抵抗と捕集ダスト量の関係
  • 粉じん容量
  • Internal production comparison
  • Formal customer test report

A system configured only for efficiency and resistance testing cannot perform complete conditioning and dust-loading procedures without the corresponding modules.

1. Airflow Test Duct

The test duct provides controlled airflow through the filter.

Its configuration depends on:

  • Maximum filter dimensions
  • Filter type
  • Rated airflow
  • Required pressure range
  • Test orientation
  • Available laboratory space

The duct should provide stable airflow, suitable flow distribution and low system leakage. The fan must overcome the resistance of the duct, filter and measurement components throughout the required operating range.

Airflow measurement and pressure sensors must cover both clean-filter and loaded-filter conditions.

2. Aerosol Generation and Mixing

Fractional-efficiency testing requires stable test aerosols over the required particle-size range.

The aerosol system must provide:

  • Suitable particle-size distribution
  • Stable concentration
  • Adequate output at maximum airflow
  • Effective mixing before upstream sampling
  • Controlled switching and cleaning

Generator quantity and configuration cannot be selected from filter size alone. Maximum airflow and required concentration are also important.

3. Particle Measurement and Sampling

Upstream and downstream particle concentrations are measured across the required size range.

The system may require:

  • Particle counters or particle-sizing instruments
  • Upstream and downstream sampling probes
  • Automatic sampling switching
  • Dilution for higher upstream concentrations
  • Controlled sampling tubing
  • Background monitoring

Instrument range, resolution and counting statistics affect the efficiency result. A particle counter with the wrong concentration range cannot be corrected simply through software.

4. Filter-Conditioning Cabinet

ISO 16890 conditioning is performed in a dedicated cabinet according to ISO 16890-4.

Its purpose is to determine the minimum fractional test efficiency after reducing the influence of electrostatic charge. It is not a dust-loading process and should not be described as field-ageing simulation.

The conditioning cabinet is a separate functional module from the main test duct.

5. Dust-Loading and Weighing Equipment

ISO 16890-3 evaluates gravimetric efficiency and resistance versus captured dust mass.

The required equipment can include:

  • Test-dust feeder
  • Dust dispersion and injection system
  • Dust preparation and handling area
  • 精密計量装置
  • Resistance monitoring
  • Exhaust or dust-collection arrangements
  • Software for loading-stage data

Dust loading is separate from the ePM classification procedure. It provides comparative loading data rather than an exact prediction of field replacement time.

6. 制御ソフトウェアとレポート作成

Software can coordinate:

  • エアフロー制御
  • エアロゾルの生成
  • Instrument switching
  • Dilution settings
  • Test sequences
  • Data acquisition
  • Efficiency calculation
  • 分類
  • Alarm handling
  • レポートの生成

Automation improves repeatability and traceability, but the calculation logic and measurement chain must still be validated.

Can One System Support Other Standards?

Some hardware may be shared between ISO 16890, ASHRAE 52.2 and legacy EN 779 テスト, 含む:

  • Test duct
  • Fan
  • Airflow measurement
  • Pressure measurement
  • Part of the particle-sampling system

しかし, the standards use different aerosols, conditioning or loading procedures, calculations and report formats.

A system should not be described as compliant with several standards unless the required hardware, instruments and software procedures for each method have been verified.

Information Required for Equipment Selection

An equipment supplier should request:

  1. Maximum filter size and construction
  2. Minimum and maximum airflow
  3. Maximum expected pressure drop
  4. Required parts of ISO 16890
  5. Target ePM range
  6. Required aerosol and instruments
  7. Test volume and automation level
  8. Laboratory dimensions and utilities
  9. Calibration and traceability requirements
  10. Future ASHRAE 52.2 または EN 779 compatibility

SCPUR configures ISO 16890 laboratories according to the filter, airflow range and required test scope. Fractional efficiency, conditioning and dust loading are treated as connected but separate capabilities, allowing manufacturers to select the modules they actually need.

ISO 16890

連絡する

出発前に...

自分のプロジェクトにどの国際標準が適用されるかわからない? 当社の 10 年間の業界専門知識に基づいた個別の技術評価については、当社の上級エンジニアと相談してください。.

*私たちはあなたの秘密を尊重し、すべての情報が保護されます.

世界標準の試験システムで品質を検証

当社の専門家がお客様に合わせてカスタム ソリューションを設計いたします. あなたのプロジェクトについて教えてください, あなたの品質を正確かつ迅速に検証しましょう.

カスタムR&D & カスタマイズされたソリューション

カスタムR&D & カスタマイズされたソリューション

世界標準への準拠

世界標準への準拠

100% 自社開発のソフトウェアシステム

100% 自社開発ソフトウェアシステム

リモート校正 & グローバルサポート

リモート校正 & グローバルサポート

お問い合わせ

*私たちはあなたの秘密を尊重し、すべての情報が保護されます.

This page has been automatically translated from the English original for convenience. Product models, standards, numerical values and technical limits should be verified against the English version. Please contact SCPUR for project-specific confirmation.