ISO 16890 테스트 장비: What Does a Complete System Include?

공유하다:

목차

ISO 16890 test equipment” does not refer to one instrument.

A complete laboratory can include fractional-efficiency measurement, 기류 저항, filter conditioning, 먼지 로딩, weighing and data-processing functions. These functions may use connected modules, but they do not all perform the same test.

The correct configuration depends on whether the laboratory needs product development, routine quality control or complete testing according to the ISO 16890 시리즈.

Define the Test Scope First

장비를 선택하기 전에, determine which results are required:

  • Initial fractional-efficiency curve
  • Conditioned fractional efficiency
  • ISO ePM classification
  • Airflow–resistance curve
  • 중량 효율성
  • Resistance versus captured dust mass
  • 먼지 용량
  • Internal production comparison
  • Formal customer test report

A system configured only for efficiency and resistance testing cannot perform complete conditioning and dust-loading procedures without the corresponding modules.

1. Airflow Test Duct

The test duct provides controlled airflow through the filter.

Its configuration depends on:

  • Maximum filter dimensions
  • Filter type
  • Rated airflow
  • Required pressure range
  • 시험 방향
  • Available laboratory space

The duct should provide stable airflow, suitable flow distribution and low system leakage. The fan must overcome the resistance of the duct, filter and measurement components throughout the required operating range.

Airflow measurement and pressure sensors must cover both clean-filter and loaded-filter conditions.

2. Aerosol Generation and Mixing

Fractional-efficiency testing requires stable test aerosols over the required particle-size range.

The aerosol system must provide:

  • Suitable particle-size distribution
  • Stable concentration
  • Adequate output at maximum airflow
  • Effective mixing before upstream sampling
  • Controlled switching and cleaning

Generator quantity and configuration cannot be selected from filter size alone. Maximum airflow and required concentration are also important.

3. Particle Measurement and Sampling

Upstream and downstream particle concentrations are measured across the required size range.

The system may require:

  • Particle counters or particle-sizing instruments
  • Upstream and downstream sampling probes
  • Automatic sampling switching
  • Dilution for higher upstream concentrations
  • Controlled sampling tubing
  • Background monitoring

Instrument range, resolution and counting statistics affect the efficiency result. A particle counter with the wrong concentration range cannot be corrected simply through software.

4. Filter-Conditioning Cabinet

ISO 16890 conditioning is performed in a dedicated cabinet according to ISO 16890-4.

Its purpose is to determine the minimum fractional test efficiency after reducing the influence of electrostatic charge. It is not a dust-loading process and should not be described as field-ageing simulation.

The conditioning cabinet is a separate functional module from the main test duct.

5. Dust-Loading and Weighing Equipment

ISO 16890-3 evaluates gravimetric efficiency and resistance versus captured dust mass.

The required equipment can include:

  • Test-dust feeder
  • Dust dispersion and injection system
  • Dust preparation and handling area
  • 정밀 계량 장비
  • Resistance monitoring
  • Exhaust or dust-collection arrangements
  • Software for loading-stage data

Dust loading is separate from the ePM classification procedure. It provides comparative loading data rather than an exact prediction of field replacement time.

6. 제어 소프트웨어 및 보고

Software can coordinate:

  • 공기 흐름 제어
  • 에어로졸 생성
  • Instrument switching
  • Dilution settings
  • Test sequences
  • Data acquisition
  • Efficiency calculation
  • 분류
  • Alarm handling
  • 보고서 생성

Automation improves repeatability and traceability, but the calculation logic and measurement chain must still be validated.

Can One System Support Other Standards?

Some hardware may be shared between ISO 16890, ASHRAE 52.2 and legacy EN 779 테스트, 포함:

  • Test duct
  • Fan
  • Airflow measurement
  • Pressure measurement
  • Part of the particle-sampling system

하지만, the standards use different aerosols, conditioning or loading procedures, calculations and report formats.

A system should not be described as compliant with several standards unless the required hardware, instruments and software procedures for each method have been verified.

Information Required for Equipment Selection

An equipment supplier should request:

  1. Maximum filter size and construction
  2. Minimum and maximum airflow
  3. Maximum expected pressure drop
  4. Required parts of ISO 16890
  5. Target ePM range
  6. Required aerosol and instruments
  7. Test volume and automation level
  8. Laboratory dimensions and utilities
  9. Calibration and traceability requirements
  10. Future ASHRAE 52.2 또는 EN 779 compatibility

SCPUR configures ISO 16890 laboratories according to the filter, airflow range and required test scope. Fractional efficiency, conditioning and dust loading are treated as connected but separate capabilities, allowing manufacturers to select the modules they actually need.

ISO 16890

연락하세요

가기 전에...

귀하의 프로젝트에 어떤 국제 표준이 적용되는지 확실하지 않음? 당사의 10년 업계 전문 지식을 바탕으로 한 맞춤형 기술 평가를 위해 수석 엔지니어와 상담하세요..

502 나쁜 게이트웨이

502 나쁜 게이트웨이


nginx
*우리는 귀하의 기밀성을 존중하며 모든 정보는 보호됩니다.

글로벌 표준 테스트 시스템으로 품질을 검증하세요

우리 전문가들은 귀하를 위한 맞춤형 솔루션을 설계할 준비가 되어 있습니다.. 귀하의 프로젝트에 대해 알려주십시오., 정확하고 빠르게 품질을 검증해 드립니다..

커스텀 R&디 & 맞춤형 솔루션

커스텀 R&디 & 맞춤형 솔루션

글로벌 스탠다드 준수

글로벌 스탠다드 준수

100% 자체 개발 소프트웨어 시스템

100% 자체 개발 소프트웨어 시스템

원격 교정 & 글로벌 지원

원격 교정 & 글로벌 지원

문의하기

*우리는 귀하의 기밀성을 존중하며 모든 정보는 보호됩니다.

This page has been automatically translated from the English original for convenience. Product models, standards, numerical values and technical limits should be verified against the English version. Please contact SCPUR for project-specific confirmation.